This study states that the Peripheral artery disease (PAD) is underdiagnosed in primary care settings, partly because of limited accuracy of the Doppler ankle-brachial index (ABI). This study aimed to assess the diagnostic accuracy of an automated oscillometric ABI device compared with reference standard vascular laboratory Doppler ABI equipment and to examine the influence of oscillometric errors on performance.
Consecutive patients who were referred to a large general hospital for an ABI measurement were invited to participate. In each patient, the oscillometric analysis was followed by the Doppler analysis. Legs with incompressible ankle arteries were excluded from analysis. ABI values were compared using the Bland-Altman method. Oscillometric errors were defined as the incapacity of the oscillometer to report a value of ABI. A receiver operating characteristic curve was constructed, and the area under the curve was computed. A total of 201 patients participated. The Bland-Altman plot showed a mean difference of 0.05 ± 0.12 (limits of agreement, −0.20 to 0.29), representing a small ABI overestimation after oscillometry. Oscillometric errors occurred more commonly in limbs with PAD than in limbs without PAD (28% and 7%, respectively; P < .001).
Reference link- https://www.jvascsurg.org/article/S0741-5214(20)31570-6/fulltext