This study states that Fenestrated-branched endovascular aneurysm repair (F/B-EVAR) is a complex procedure that generates high radiation doses. Magnification aids in vessel cannulation but increases radiation. The aim of the study was to compare radiation doses to patients and operating room staff from two fluoroscopy techniques, standard magnification vs dual fluoroscopy with live-image digital zooming during F/B-EVAR.

An observational, prospective, single-center study of F/B-EVAR procedures using Philips Allura XperFD20 equipment (Philips Healthcare, Amsterdam, The Netherlands) was performed during a 42-month period. Intravascular ultrasound, three-dimensional fusion, and extreme collimation were used in all procedures. Intraoperative live-image processing was performed with two imaging systems: standard magnification in 123 patients (81%) and dual fluoroscopy with live-image digital zooming in 28 patients (18%). In the latter, the live “processed” zoomed images are displayed on examination displays and live images are displayed on reference displays. The reference air kerma was collected for each case and represents patient dose. Operating staff personal dosimetry was collected using the DoseAware system (Philips Healthcare). Patient and staff radiation doses were compared using nonparametric tests.

Reference link- https://www.jvascsurg.org/article/S0741-5214(20)31286-6/fulltext